Overview
Fluorescence Staining & Latex-enhanced immunonephelometric assay Diagnostics management of Common Fever Diseases
The DX-80 series provides high quality test results, automation operation , three-in-one detection (CBC,CRP,SAA) and stable use experience. In terms of methodology, the advanced semiconductor laser flow cytometry + nucleic acid fluorescence staining three-dimensional analysis technology is adopted, and the latex enhanced immunoscattering turbidity principle technology is used.Only 20µl of peripheral blood is required for precise analysis of blood diseases, enabling early warning of febrile illnesses.
80T/H
Throughput
20ul
Capillary blood
29ul
CBC+CRP+SAA
Automatic
workcell
Advantage
Flow cytometry
& Fluorescent staining method
Automatic sampling
with 20ul Capillary blood
CBC & DIFF
& CRP & SAA in one tube
Automatically retest
by user customized rules
Major Parameters

Analysis Principle:

DC impedance method:WBC/RBC/PLT
Colorimetric method:HGB
FCM & Fluorescence Staining:DIFF
Latex-enhanced immunonephelometric assay:CRP/SAA

Report Parameters:

Hematology: 26 reportable parameters and 5 research parameters
3 histograms (WBC/RBC/PLT); 1 scatter grams(DIFF)
CRP test: 3 reportable parameters(CRP/FR-CRP/hs-CRP)
SAA test: 1 reportable parameter(SAA)

 

   
Measurement Mode DX-80 CBC/CBC+DIFF
DX-80C CBC/CBC+DIFF/CBC+CRP/CBC+DIFF+CRP
DX-80S CBC/CBC+DIFF/CBC+CRP/CBC+DIFF+CRP/CBC+DIFF+CRP+SAA…
Sample Type Venial blood,Peripheral blood,Pre-dilution blood,Serum
Throughput  CBC+DIFF: 80T/H CBC+DIFF+CRP: 60T/H
Sample Loading Manual,Auto-sampler for normal tube with cap,Automatic mixing of micro-tube
Linearity CRP: 0.5-320mg/L,covering the full range of CRP detection
Dimensions 662(W)*870(D)*658(H)mm,≤79kg