Overview
Fluorescence Staining & Latex-enhanced immunonephelometric assay technology for laboratory testing related to common febrile conditions.
The DX-80 series delivers consistent test results with automated operation, supporting three-in-one detection (CBC, CRP, SAA). It employs semiconductor laser flow cytometry combined with nucleic acid fluorescence staining three-dimensional analysis, alongside latex-enhanced immunoscatter turbidimetry technology.
A peripheral blood sample volume of 20µL enables precise analysis of hematological parameters to support laboratory applications related to febrile illnesses.
80T/H
Throughput
20ul
Capillary blood
29ul
CBC+CRP+SAA
Automatic
workcell
Advantage
Flow cytometry
& fluorescent staining technology
Automatic sampling
with 20µL capillary blood
Simultaneous detection of CBC,
DIFF, CRP and SAA in one tube
Automatic retest
function configurable by user-defined rules
Major Parameters

Analysis Principle:

DC impedance method:WBC/RBC/PLT
Colorimetric method:HGB
FCM & Fluorescence Staining:DIFF
Latex-enhanced immunonephelometric assay:CRP/SAA

Report Parameters:

Hematology: 26 reportable parameters and 5 research parameters
3 histograms (WBC/RBC/PLT); 1 scatter grams(DIFF)
CRP test: 3 reportable parameters(CRP/FR-CRP/hs-CRP)
SAA test: 1 reportable parameter(SAA)

 

   
Measurement Mode DX-80 CBC/CBC+DIFF
DX-80C CBC/CBC+DIFF/CBC+CRP/CBC+DIFF+CRP
DX-80S CBC/CBC+DIFF/CBC+CRP/CBC+DIFF+CRP/CBC+DIFF+CRP+SAA…
Sample Type Venial blood,Peripheral blood,Pre-dilution blood,Serum
Throughput  CBC+DIFF: 80T/H CBC+DIFF+CRP: 60T/H
Sample Loading Manual,Auto-sampler for normal tube with cap,Automatic mixing of micro-tube
Linearity CRP: 0.5-320mg/L
Dimensions 662(W)*870(D)*658(H)mm,≤79kg